If your ultrasonic cleaner is working effectively the foil should be perforated.
Ultrasonic cleaner test log.
Turn off the ultrasonic cleaner after 60 seconds.
It is invisible to the human eye but with the sonocheck you can test for the presence of cavitation energy the cleaning power of your ultra sonic cleaner.
Use the sonic cleaner test log sheet to record test results.
Then switch the unit on.
Remove the foil from the tank and allow it to dry.
Prepare a fresh tank of the cleaning solution.
Problems such as insufficient energy water level improper temperature and degassing may have an impact on the results.
The verify ultrasonic indicator provides an independent objective test to evaluate the ultrasonic cycle.
Think of this as akin to a service record for your new car or your medical records kept on file by your family physician.
To test the working fill in the ultrasonic cleaner tank with water and ultrasonic cleaning solution.
Log book form.
Ultrasonic cleaning is a result of sound waves introduced into a cleaning liquid by a series of transducers.
By just observing the activity in the tank for 2 3 minutes you can easily tell whether your ultrasonic cleaning machine is working or not.
If it s working there can be some activity seen as a ripple effect on the surface of the solution.
Date cleaning disinfecting of bath solution filled in am solution emptied in pm extra solution change strip test strip test results pass or fail initials please contact your maxill representative for any questions 1 800 268 8633.
Turn on the cleaner and run it on high temperature.
Generalsonic gs3 foil test below is the result of a foil test carried out with a generalsonic gs3.
Log sheet ultrasonic cleaner test machine number.
How to set an ultrasonic cleaner performance benchmark.
Ultrasonic cleaner details this logbook covers 1 year beginning.
To do a foil test do the following steps.
Cleaning solution changing test log.
Ultrasonic cleaner uslog.
Turn the unit on and de gas accord ing to the manufactures instructions 3.
Place the wire frame with the aluminium foil into the ultrasonic cleaner.
To know if an ultrasonic cleaner is cavitating properly a foil test can be done.
Record the date tester initials sonocheck result left side center right side tank degassed bath temp program cycle tested action comments.
Normal cycle time facility name date started quick steps for testing sonic cavita tion using the sonocheck tm 1.
Ultrasonic cleaning uses cavitation chemistry time and temperature to effectively remove soils after manual cleaning.
Include the equipment name and model of the ultrasonic washer being tested.
A widely accepted means to establish an ultrasonic cleaner performance benchmark is called the foil test it is simple.